User Tools

Site Tools


slice:dcc:api

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
Next revisionBoth sides next revision
slice:dcc:api [2019/12/10 05:30] – [Global SLICE Commands] Michael Radunskyslice:dcc:api [2019/12/11 02:59] – [SLICE-DCC-specific Commands] Michael Radunsky
Line 106: Line 106:
 __Possible system responses__\\ __Possible system responses__\\
 Requested current in Amps or closest value as bounded by unit capacity.\\ Requested current in Amps or closest value as bounded by unit capacity.\\
 +<WRAP center round box 60%>
 +| {{ :take-note.png?200&nolink }} | CURRSET does not turn on or off the current. It only changes the set point value. Use a CONTROL command to change the operation mode. |
 +</WRAP>
 ---- ----
 <function=ICEfunc|0=MAXCURR?&1=I:CHANNEL:2&2=1.435300&-1> <function=ICEfunc|0=MAXCURR?&1=I:CHANNEL:2&2=1.435300&-1>
-Returns user-set maximum current in Amps for channel CHANNEL. In the example, the maximum current of channel 2 has been set to 1.435300 A.\\ \\+Returns user-set maximum current in Amps for channel CHANNEL. In the example, the maximum current supplied by channel 2 has been set to 1.435300 A.\\ \\
 __Input range__\\ __Input range__\\
 CHANNEL: {1,2}\\ CHANNEL: {1,2}\\
Line 134: Line 137:
 CHANNEL: {1,2}\\ CHANNEL: {1,2}\\
 ---- ----
-<function=ICEfunc|0=PWRSET&1=I:CHANNEL:2;F:POWER:314.0&2=1.3&-1>+<function=ICEfunc|0=PWRSET&1=I:CHANNEL:2;F:POWER:314.0&2=314.0&-1>
 Sets the set point power level in milliwatts to which the Constant Power Mode will attempt to servo for channel CHANNEL. Returns PWRSET? CHANNEL. In the example, the set point power for channel 2 is set to 314.0 mW.\\ \\ Sets the set point power level in milliwatts to which the Constant Power Mode will attempt to servo for channel CHANNEL. Returns PWRSET? CHANNEL. In the example, the set point power for channel 2 is set to 314.0 mW.\\ \\
 __Input parameter ranges__\\ __Input parameter ranges__\\
Line 153: Line 156:
 ---- ----
 <function=ICEfunc|0=GAIN?&1=I:CHANNEL:1&2=30.000000&-1> <function=ICEfunc|0=GAIN?&1=I:CHANNEL:1&2=30.000000&-1>
-Returns the Integral Gain of the Constant Power feedback loop for channel CHANNEL in dB. In the example, the Gain for Constant Power mode feedback for channel 1 is 30.000000 dB. The default value of the Gain is +30 dB.\\ \\+Returns the Integral Gain of the Constant Power feedback loop for channel CHANNEL in dB. In the example, the Gain for Constant Power mode feedback for channel 1 is +30.000 000 dB. The default value of the Gain is +30.000 000 dB.\\ \\
 __Input range__\\ __Input range__\\
 CHANNEL: {1,2}\\ CHANNEL: {1,2}\\
Line 168: Line 171:
 CHANNEL: {1,2}\\ CHANNEL: {1,2}\\
 ---- ----
-<function=ICEfunc|0=RESPVTY&1=I:CHANNEL:2;F:GAIN:0.0035&2=0.0035&-1>+<function=ICEfunc|0=RESPVTY&1=I:CHANNEL:1;F:RESPONSIVITY:0.0035&2=0.0035&-1>
 Sets the Constant Power mode external (customer-supplied) detector responsivity in Amps per Watt for channel CHANNEL. Returns RESPVTY? CHANNEL. In the example, the responsivity of the detector for channel 1 is set to 0.0035 A/W.\\ \\ Sets the Constant Power mode external (customer-supplied) detector responsivity in Amps per Watt for channel CHANNEL. Returns RESPVTY? CHANNEL. In the example, the responsivity of the detector for channel 1 is set to 0.0035 A/W.\\ \\
 __Input parameter ranges__\\ __Input parameter ranges__\\
 CHANNEL: {1,2}\\ CHANNEL: {1,2}\\
-GAIN: (TBD)\\+RESPONSIVITY: (TBD)\\
 ---- ----
 <function=ICEfunc|0=POL?&1=I:CHANNEL:1&2=OFF&-1> <function=ICEfunc|0=POL?&1=I:CHANNEL:1&2=OFF&-1>
Line 198: Line 201:
 ---- ----
 <function=ICEfunc|0=AMODSEL?&1=I:CHANNEL:1&2=1&-1> <function=ICEfunc|0=AMODSEL?&1=I:CHANNEL:1&2=1&-1>
-Returns the analog modulation source for channel CHANNEL. In the example, the modulation source for channel 1 is the front-panel input.\\ \\+Returns the analog modulation source for channel CHANNEL. In the example, the modulation source for channel 1 is the front-panel input (labeled A).\\ \\
 __Return Values__\\ __Return Values__\\
 0 = Modulation input is the SMA connector labeled Mod In on the back of the unit for the respective channel (default).\\ 0 = Modulation input is the SMA connector labeled Mod In on the back of the unit for the respective channel (default).\\
Line 299: Line 302:
 CHANNEL: {1,2}\\ CHANNEL: {1,2}\\
 __Possible output values__\\ __Possible output values__\\
-Power: (0 to 41.5 W)+Power: (0 to 42.5 W)
 ---- ----
 ===== Error Handling ===== ===== Error Handling =====
slice/dcc/api.txt · Last modified: 2023/12/06 20:35 by Christiana Sasser