slice:z-n
This is an old revision of the document!
Ziegler-Nichols Loop Tuning
Warning: There is the possibility for the loop to go unstable. Use caution as gain is increased.
- From the Home Screen, touch the Channel icon for the loop to be tuned
- Touch Settings > PID Params
- Choose a Set Point temperature
- Set Proportional Gain (GP) to 0
- Turn off Integral and Differential portions of the loop (TI & TD, respectively)
- While monitoring the error in temperature on rolling graph, slowly increase GP until the error begins to oscillate.
- Use the Cursors enabled from the
slice/z-n.1554217021.txt.gz · Last modified: 2021/08/26 14:26 (external edit)