ice:commands:lasercurrent
Common Laser Controller Command Set
List of common commands for controlling the laser controller in both the Peak Lock Servo and Peak Lock Servo & Current Controller and the Offset Phase Lock Servo & Current Controller.
Laser
Arguments:
No Arguments Taken
Example:
Laser
I2C Command Number:
Description
Returns the status of the laser (on, off or fault). <html></div></html>
Laser
Arguments:
[ASCII] On/Off
Example:
Laser On Fault
I2C Command Number: 17
Description
Turns the laser on or off. Valid arguments are ON or OFF. Only turns the laser on when no fault / interlock issues. Returns the output from the Laser? command. <html></div></html>
CurrSet
Arguments:
No Arguments Taken
Example:
CurrSet
I2C Command Number:
Description
Returns the laser current setpoint (in mA) for how much current will flow through the laser when the laser is on. <html></div></html>
CurrSet
Arguments:
[Float] CURRENT
Example:
CurrSet 123.52 123.5
I2C Command Number: 19
Description
Sets the setpoint laser current (in mA) to CURRENT. Returns the output from the CurrSet? command. <html></div></html>
CurrLim
Arguments:
No Arguments Taken
Example:
CurrLim
I2C Command Number:
Description
Returns the current limit (in mA) for the maximum current allowed to flow through the laser when the laser is on. <html></div></html>
CurrLim
Arguments:
[Float] CURRENT
Example:
CurrLim 156 156
I2C Command Number: 21
Description
Sets the maximum laser current to CURRENT. Returns the output from the CurrLim? command. Note: If Laser Current is set above new current limit value, then current setpoint is lowered to new current limit value. <html></div></html>ice/commands/lasercurrent.txt · Last modified: 2023/11/30 21:29 by 127.0.0.1