Previous revision
d2:dbr_laser [2022/08/17 21:56] – external edit 127.0.0.1d2:dbr_laser [2024/03/29 22:18] (current) – [Absolute Maximum Ratings] Thomas Bersano
Line 48: Line 48:
 ===== Absolute Maximum Ratings ===== ===== Absolute Maximum Ratings =====
  
-Note: All modules designed to be operated in laboratory environment+Note: All modules designed to be operated in laboratory environment.
 <WRAP center round box 60%> <WRAP center round box 60%>
 ^ Parameter ^ Rating ^ ^ Parameter ^ Rating ^
Line 130: Line 130:
  
 ==== Safety Information ==== ==== Safety Information ====
 +<WRAP center round important 60%>
 +The D2-100 is only specified for use with the Vescent D2-105 laser controller, and with cables provided by Vescent.
 +</WRAP>
 +
 The D2-100 is a Class 3B CW laser product under the guidelines provided in IEC 60825-01:2014. It has no serviceable parts, and should not be disassembled in the field under any circumstance. Viewing of the beam, either directly, with viewing optics, or from a specular reflection can cause damage to the eye, and should be avoided.  The D2-100 is a Class 3B CW laser product under the guidelines provided in IEC 60825-01:2014. It has no serviceable parts, and should not be disassembled in the field under any circumstance. Viewing of the beam, either directly, with viewing optics, or from a specular reflection can cause damage to the eye, and should be avoided. 
  
Line 136: Line 140:
 The Maximum Permissable Exposire (MPE) for the D2-100 is 14.45 W*m<sup>-2</sup>, calculated with a >10s exposure time, and the Nominal Ocular Hazard distance is roughly 100m. Proper eye protection for the D2-100 is highly wavelength dependent. We recommend [[https://noirlaser.com/lasershields?input1=780&input2=0&input3=0|this]] online tool, provided by NoIR, and an OD value of at least 3+ for the selected wavelength when choosing the correct eye protection. The Maximum Permissable Exposire (MPE) for the D2-100 is 14.45 W*m<sup>-2</sup>, calculated with a >10s exposure time, and the Nominal Ocular Hazard distance is roughly 100m. Proper eye protection for the D2-100 is highly wavelength dependent. We recommend [[https://noirlaser.com/lasershields?input1=780&input2=0&input3=0|this]] online tool, provided by NoIR, and an OD value of at least 3+ for the selected wavelength when choosing the correct eye protection.
  
-There are three safety stickers present on the D2-100. The first, located on the side of the isolator module, is the explanatory sticker, which states the classification of the laser, and several safety instructions. It reads: "Invisible laser radiation avoid exposure to beam class 3B laser product".+There are four safety stickers present on the D2-100. The first, located on the side of the isolator module, is the explanatory sticker, which states the classification of the laser, and several safety instructions. It reads: "Invisible laser radiation avoid exposure to beam class 3B laser product".
  
 <imgcaption safety_sticker|Reproduction of Explanatory label on the D2-100. \>{{ :d2:d2-100:sticker.png?200 |}}</imgcaption> <imgcaption safety_sticker|Reproduction of Explanatory label on the D2-100. \>{{ :d2:d2-100:sticker.png?200 |}}</imgcaption>
 +
 +On the side of the isolator module opposite the explanatory sticker is a sticker which lists the maximum possible output of the laser in mW, the wavelength of the laser, and the name of the IEC certification standard to which the laser adheres.
 +
 +<imgcaption IEC_sticker|Reproduction of second Explanatory label on the D2-100.>{{ :d2:d2-100:needed_sticker.png?200 |}}</imgcaption>
  
 On the top of the D2-100, near the laser aperture, is a "Laser Aperture" sticker. On the top of the D2-100, near the laser aperture, is a "Laser Aperture" sticker.
Line 149: Line 157:
  
 There is only one laser aperture on the D2-100, located on the front with the isolator as shown in (<imgref DBR_laser>).  There is only one laser aperture on the D2-100, located on the front with the isolator as shown in (<imgref DBR_laser>). 
 +
 +Safe operation of the D2-100 necessitates that a compliant laser driver be used to supply a current to the diode. It is important to always correctly set the current limit on any laser driver before energizing the laser system. The D2-100 is a Cathode-Ground laser, and requires a DC current between 0mA and 500mA((Specific current for each laser supplied with test documentation)) to be safely turned on. The connector used to supply current to the D2-100 is a female SMA, which requires a male SMA cable for compatibility.
d2/dbr_laser.1660773373.txt.gz · Last modified: 2022/08/17 21:56 by 127.0.0.1