slice:qt [2019/06/12 00:52] – [System Settings] Michael Radunsky | slice:qt [2019/08/22 22:40] – [Home Screen] Michael Radunsky |
---|
| **Parameter** | **Value** | | | **Parameter** | **Value** | |
| Channels | 4 | | | Channels | 4 | |
| Loop FIlter | PID, adjustable corners and proportional Gain | | | Loop Filter | PID, adjustable corners and proportional Gain | |
| Control Range((Lower precision may be realized above 50°C. Consider using a thermistor with 10 kΩ at higher temperature than nominal.)) | -10 to +120°C | | | Control Range((Lower precision may be realized above 50°C. Consider using a thermistor with 10 kΩ at higher temperature than nominal.)) | -10 to +120°C | |
| Compatible Transducer | TEC or resistive heater | | | Compatible Transducer | TEC or resistive heater | |
| Compatible Sensor | NTC thermistor ((Nominal value 10 kΩ at room temperature. Precision at temperatures removed from room temperature can be improved by selecting a thermistor whose resistance is ~10 kΩ near the target temperature.)) | | | Compatible Sensor | NTC thermistor ((Nominal value 10 kΩ at room temperature. Precision at temperatures removed from room temperature can be improved by selecting a thermistor whose resistance is ~10 kΩ near the target temperature.)) | |
| Temperature Stability | ±0.2 mK over 1 hour((Plant-dependent; value specified for D2-100 laser)) | | | Temperature Stability | ±0.2 mK over 1 hour((Plant-dependent; value specified for D2-100 laser)) | |
| Precision | ~1 mK | | | Precision | ~1 mK | |
| Control Capacity | 35 W((User-distributable over four channels)) | | | Control Capacity | 35 W((User-distributable over four channels)) | |
| Current Capacity | 6 A per channel((Subject to 35 W overall maximum)) | | | Current Capacity | 6 A per channel((Subject to 35 W overall maximum)) | |
In most instances, there are two methods for entering parameter values: the physical rotary knobs on the front panel and the pop-up keypad on the touch screen. When using the knobs to change the value of a parameter, changes take effect immediately. When using the keypad, changes do not take effect until the enter key is depressed. | In most instances, there are two methods for entering parameter values: the physical rotary knobs on the front panel and the pop-up keypad on the touch screen. When using the knobs to change the value of a parameter, changes take effect immediately. When using the keypad, changes do not take effect until the enter key is depressed. |
=== Rotary Knobs === | === Rotary Knobs === |
If you touch an editable field for a brief moment, a cursor will appear under one of the digits of the parameter value. By turning the right knob, it is possible to increase (cw) or decrease (ccw) the value of the underlined digit. Turning the left knob will change which digit is editable (cw to move to the left and ccw to move to the right). | If you touch an editable field for a brief moment, a cursor will appear under one of the digits of the parameter value. By turning the right knob, it is possible to increase (cw) or decrease (ccw) the value of the underlined digit. Turning the left knob will change which digit is editable (cw to move to the right and ccw to move to the left). |
== Keypad == | == Keypad == |
If you touch and hold an editable field such as the set point temperature, a numeric keypad will appear as seen in <imgref keypad>. You can enter in the full value of the desired parameter with this keypad.\\ | If you touch and hold an editable field such as the set point temperature, a numeric keypad will appear as seen in <imgref keypad>. You can enter in the full value of the desired parameter with this keypad.\\ |
To set the temperature set point from the home screen, touch the Setpoint window for the desired channel (circled in red for CH 1 in <imgref home_Tset1>). A long touch will cause a keypad to appear. Enter desired value and touch the accept button. | To set the temperature set point from the home screen, touch the Setpoint window for the desired channel (circled in red for CH 1 in <imgref home_Tset1>). A long touch will cause a keypad to appear. Enter desired value and touch the accept button. |
| |
A short touch will highlight the Setpoint window in yellow (<imgref home_Tset2>). Use rotary knobs to select digit to edit (right knob) and value for digit (left knob).\\ | A short touch will highlight the Setpoint window in yellow (<imgref home_Tset2>). Use rotary knobs to select digit to edit (left knob) and value for digit (right knob).\\ |
<imgcaption home_Tset1|Select Temperature setpoint for CH 1>{{ :slice:slice-qt_set_t_home_1.png?400x254 |}}</imgcaption> | <imgcaption home_Tset1|Select Temperature setpoint for CH 1>{{ :slice:slice-qt_set_t_home_1.png?400x254 |}}</imgcaption> |
| |