=====Description===== | =====Description===== | ||
The Power Supply is a quiet, linear power supply providing +/-15V and 5V. It is in a separate box to isolate 60 Hz and EM interference. One power supply can power two complete laser systems, or four electronics modules. | The Power Supply is a quiet, linear power supply providing +/-15V and 5V. It is in a separate box to isolate 60 Hz and EM interference. One power supply can power two complete laser systems, or four electronics modules. | ||
+ | |||
+ | =====Symbols===== | ||
+ | A list of symbols which appear on the D2-005, and their definitions, | ||
+ | |||
+ | {{: | ||
+ | |||
+ | {{: | ||
+ | |||
+ | {{: | ||
+ | |||
+ | {{: | ||
+ | |||
+ | |||
+ | |||
+ | =====Warnings===== | ||
+ | This manual must be consulted in all cases where the Caution symbol is marked, in order to find out the nature of the potential hazards and any actions which have to be taken to avoid them. | ||
+ | |||
+ | **Caution: Risk of Electrical Shock Do Not Open** -- The D2-005 contains high voltage components and should never be opened in any circumstance except by trained Vescent personnel. | ||
+ | |||
+ | <WRAP center round important 60%> | ||
+ | There are no serviceable parts in the D2-005. Only Vescent is authorized to service this product. | ||
+ | </ | ||
+ | <WRAP center round important 60%> | ||
+ | Nothing must obstruct airflow from the top panel vent. Never place anything on top of the D2-005. | ||
+ | </ | ||
+ | <WRAP center round important 60%> | ||
+ | A potential hazard exists if the D2-005 comes into contact with water or other conductive liquids, and exposure to these should be avoided. | ||
+ | </ | ||
+ | |||
+ | Special care must be given to replacing the provided mains power cord, such that any replacement cords are appropriately rated. | ||
+ | |||
+ | If the D2-005 is used in a manner not specified by Vescent or in this manual, the protection provided by the equipment may be impaired. | ||
+ | =====Purchase Includes===== | ||
+ | * D2-005 Power Supply | ||
+ | * VPN00464 Power Cord | ||
+ | |||
+ | |||
+ | |||
===== Absolute Maximum Ratings ===== | ===== Absolute Maximum Ratings ===== | ||
- | Note: All modules designed to be operated in laboratory environment | + | Note: All modules designed to be operated in an indoor |
<WRAP center round box 60%> | <WRAP center round box 60%> | ||
| **Parameter** | **Rating** | | | **Parameter** | **Rating** | | ||
Line 33: | Line 81: | ||
^ ^^ | ^ ^^ | ||
^Current Limit | | | | ^Current Limit | | | | ||
- | |< | + | |+15V| |
- | |< | + | |-15V| |
- | |< | + | |+5V| 6 | A | |
^Output Regulation | 0.05 | % | | ^Output Regulation | 0.05 | % | | ||
+ | ^Input Voltages | 100±10, 120±10, 230±10 | ||
+ | ^Input Frequency| | ||
+ | ^Power Rating | ||
</ | </ | ||
+ | Replaceable fuses located in the power entry module are rated at 2A, 250V @115VAC or 1A, 250V @230VAC | ||
+ | =====Intended Use===== | ||
+ | The D2-005 is intended to supply low-noise DC electrical power to the following Vescent D2 products: | ||
+ | *D2-105 Laser Controller | ||
+ | *D2-125 Reconfigurable Laser Servo | ||
+ | *D2-135 Offset Phase Lock Servo | ||
+ | *D2-210 Spectroscopy Module | ||
+ | *D2-260 Beat Note Detector | ||
=====Inputs, | =====Inputs, | ||
Line 51: | Line 109: | ||
The monitor section contains three blue status LEDs indicating proper voltage on the +15,-15, and 5V power lines. If there is a voltage sag on any of the power lines, the blue status LED will turn off. | The monitor section contains three blue status LEDs indicating proper voltage on the +15,-15, and 5V power lines. If there is a voltage sag on any of the power lines, the blue status LED will turn off. | ||
+ | ====Front Panel Controls==== | ||
+ | A single switch on the front panel of the D2-005 toggles whether power is enabled or disabled. Flipping the switch to the " | ||
+ | |||
+ | <WRAP center round important 60%> | ||
+ | Do not position the D2-005 such that it is difficult to operate the switch. | ||
+ | </ | ||
====Back panel I/O==== | ====Back panel I/O==== | ||
**Power I/O (9-pin D-sub)** | **Power I/O (9-pin D-sub)** | ||
+ | <WRAP center round important 60%> | ||
+ | Always turn off the D2-005, make connections to the device, and then re-energize the D2-005. | ||
+ | </ | ||
- | The power to each electronics | + | |
+ | Power is supplied to the D2-005 via a standard C13 to Type-B | ||
{{ : | {{ : | ||
+ | |||
+ | In some system configurations, | ||
+ | |||
+ | <WRAP center round important 60%> | ||
+ | If you are considering using an Uninterruptible Power Supply (UPS) to guarantee operation of your Vescent products through a brown out or black out, great care should be taken in choosing the model. | ||
+ | </ | ||