- | =====Aligning the DBR Laser Module===== | ||
- | The module should not need adjustments, | ||
- | |||
- | - Loosen the three locking screws on the back of the DBR module ½ turn past the crack point. | ||
- | - With a 0.050" Allen driver, adjust the vertical adjustment setscrew on the top of the DBR housing to level the beam. | ||
- | - Adjust the horizontal adjustment 4-40 set screw on the left side of the DBR housing (See <imgref DBR_lock_screws> | ||
- | - Alternatively, | ||
- | - Gently retighten the three locking screws. | ||