Line 51: Line 51:
 ===== Proper Usage ===== ===== Proper Usage =====
 | {{ :​warning-sign.png?​100&​nolink }}  | If this instrument is used in a manner not specified by the manufacturer in this manual or other relevant literature, protection provided by the instrument may be impaired. | | {{ :​warning-sign.png?​100&​nolink }}  | If this instrument is used in a manner not specified by the manufacturer in this manual or other relevant literature, protection provided by the instrument may be impaired. |
 +|:::| Successful implementation of the SLICE-QTC depends critically on the design of the whole system: controller, transducer, plant, and sensor. |
 +
 +==== Tips for Successful Temperature Regulation ==== 
 +  -  The transducer, plant, and sensor must all be in good thermal contact. Use thermal paste or epoxy to avoid air gaps between a TEC or heater and the body of the plant as well as to secure the thermistor to the plant.
 +  - Know the electrical limits of of your transducer and set the limits of the SLICE-QTC accordingly. ​ Each channel can provide up to  6 A or 20 W of power with a maximum compliance voltage up to ~18 V.  Choose transducers that work within these limits. ​ Using the current, power, and/or voltage ratings of your transducer to calculate loads on individual channels and set limits appropriately in the CH X > Settings > Load Limits window.
 +  - Maintain proper shielding of the control cable, but avoid ground loops between the SLICE-QTC and the plant. ​ In most cases this means extending the foil shield in the cable to as close as possible to the plant without electrically connecting the SLICE-QTC to the plant.  ​
  
 ===== List of Symbols ===== ===== List of Symbols =====
Line 65: Line 71:
 | **Parameter** ​                                                                                                                            | **Value** ​                                                                                                                                                                                                             | | **Parameter** ​                                                                                                                            | **Value** ​                                                                                                                                                                                                             |
 | Channels ​                                                                                                                                 | 4                                                                                                                                                                                                                      | | Channels ​                                                                                                                                 | 4                                                                                                                                                                                                                      |
-| Loop FIlter ​                                                                                                                              | PID, adjustable corners and proportional Gain                                                                                                                                                                          |+| Loop Filter ​                                                                                                                              | PID, adjustable corners and proportional Gain                                                                                                                                                                          |
 | Control Range((Lower precision may be realized above 50°C.  Consider using a thermistor with 10 kΩ at higher temperature than nominal.)) ​ | -10 to +120°C ​                                                                                                                                                                                                         | | Control Range((Lower precision may be realized above 50°C.  Consider using a thermistor with 10 kΩ at higher temperature than nominal.)) ​ | -10 to +120°C ​                                                                                                                                                                                                         |
 | Compatible Transducer ​                                                                                                                    | TEC or resistive heater ​                                                                                                                                                                                               | | Compatible Transducer ​                                                                                                                    | TEC or resistive heater ​                                                                                                                                                                                               |
 | Compatible Sensor ​                                                                                                                        | NTC thermistor ((Nominal value 10 kΩ at room temperature. Precision at temperatures removed from room temperature can be improved by selecting a thermistor whose resistance is ~10 kΩ near the target temperature.)) ​ | | Compatible Sensor ​                                                                                                                        | NTC thermistor ((Nominal value 10 kΩ at room temperature. Precision at temperatures removed from room temperature can be improved by selecting a thermistor whose resistance is ~10 kΩ near the target temperature.)) ​ |
 | Temperature Stability ​                                                                                                                    | ±0.2 mK over 1 hour((Plant-dependent;​ value specified for D2-100 laser)) ​                                                                                                                                              | | Temperature Stability ​                                                                                                                    | ±0.2 mK over 1 hour((Plant-dependent;​ value specified for D2-100 laser)) ​                                                                                                                                              |
-| Precision ​                                                                                                                                | ~1 mK                                                                                                                                                                                                                   ​+| Precision ​                                                                                                                                | ~1 mK                                                                                                                                                                                                                  
-| Control Capacity ​                                                                                                                         | 35 W((User-distributable over four channels)) ​                                                                                                                                                                         | +| Control Capacity ​                                                                                                                         | 40 total((User-distributable over four channels)) ​\\ 20 W max for a single channel ​                                                                                                                                                                         | 
-| Current Capacity ​                                                                                                                         | 6 A per channel((Subject to 35 W overall maximum)) ​                                                                                                                                                                    |+| Current Capacity ​                                                                                                                         | 6 A per channel((Subject to 40 W overall maximum)) ​                                                                                                                                                                    |
 | Compliance Voltage ​                                                                                                                       | ≤18 V((automatically adjusted)) ​                                                                                                                                                                                       | | Compliance Voltage ​                                                                                                                       | ≤18 V((automatically adjusted)) ​                                                                                                                                                                                       |
 |  **Input & Output** ​                                                                                                                                                                                                                                                                                                                                              || |  **Input & Output** ​                                                                                                                                                                                                                                                                                                                                              ||
Line 81: Line 87:
 | Connections ​                                                                                                                              | Host control: USB Type B                                                                                                                                                                                               | | Connections ​                                                                                                                              | Host control: USB Type B                                                                                                                                                                                               |
 </​tabcaption></​WRAP>​ </​tabcaption></​WRAP>​
 +
 +==== Extendable Legs ====
 +The SLICE-QTC has four extendable legs (<imgref belly>​). ​ Extending the front legs allows easy viewing from above (<imgref qt_prod_shot>​) and extending the rear legs (<imgref side>) allows easy viewing from below.
 +
 +<WRAP half column><​imgcaption belly|Underside of SLICE-QTC>​{{ :​slice:​dcc:​dcc_underbelly_labelled.png |}}</​imgcaption>​
 +</​WRAP>​
 +<WRAP half column><​imgcaption side|Side view of SLICE-QTC configured for viewing from below>{{ : slice:​dcc:​slice-dcc_looking_down.png?​400 |}}</​imgcaption></​WRAP>​
  
 ===== Front Panel ===== ===== Front Panel =====
Line 154: Line 167:
 In most instances, there are two methods for entering parameter values: the physical rotary knobs on the front panel and the pop-up keypad on the touch screen. ​ When using the knobs to change the value of a parameter, changes take effect immediately. ​ When using the keypad, changes do not take effect until the enter key is depressed. In most instances, there are two methods for entering parameter values: the physical rotary knobs on the front panel and the pop-up keypad on the touch screen. ​ When using the knobs to change the value of a parameter, changes take effect immediately. ​ When using the keypad, changes do not take effect until the enter key is depressed.
 === Rotary Knobs === === Rotary Knobs ===
-If you touch an editable field for a brief moment, a cursor will appear under one of the digits of the parameter value. ​ By turning the right knob, it is possible to increase (cw) or decrease (ccw) the value of the underlined digit. ​ Turning the left knob will change which digit is editable (cw to move to the left and ccw to move to the right).  ​+If you touch an editable field for a brief moment, a cursor will appear under one of the digits of the parameter value. ​ By turning the right knob, it is possible to increase (cw) or decrease (ccw) the value of the underlined digit. ​ Turning the left knob will change which digit is editable (cw to move to the right and ccw to move to the left).  ​
 == Keypad == == Keypad ==
 If you touch and hold an editable field such as the set point temperature,​ a numeric keypad will appear as seen in <imgref keypad>​. ​ You can enter in the full value of the desired parameter with this keypad.\\ ​ If you touch and hold an editable field such as the set point temperature,​ a numeric keypad will appear as seen in <imgref keypad>​. ​ You can enter in the full value of the desired parameter with this keypad.\\ ​
Line 177: Line 190:
 To set the temperature set point from the home screen, touch the Setpoint window for the desired channel (circled in red for CH 1 in <imgref home_Tset1>​). ​ A long touch will cause a keypad to appear. ​ Enter desired value and touch the accept button. ​ To set the temperature set point from the home screen, touch the Setpoint window for the desired channel (circled in red for CH 1 in <imgref home_Tset1>​). ​ A long touch will cause a keypad to appear. ​ Enter desired value and touch the accept button. ​
  
-A short touch will highlight the Setpoint window in yellow (<imgref home_Tset2>​). ​ Use rotary knobs to select digit to edit (right knob) and value for digit (left knob).\\+A short touch will highlight the Setpoint window in yellow (<imgref home_Tset2>​). ​ Use rotary knobs to select digit to edit (left knob) and value for digit (right knob).\\
 <​imgcaption home_Tset1|Select Temperature setpoint for CH 1>{{ :​slice:​slice-qt_set_t_home_1.png?​400x254 |}}</​imgcaption>​ <​imgcaption home_Tset1|Select Temperature setpoint for CH 1>{{ :​slice:​slice-qt_set_t_home_1.png?​400x254 |}}</​imgcaption>​
  
Line 369: Line 382:
 It is possible to control the SLICE-QTC from a Windows<​sup>​®</​sup>​-based PC using a GUI.  In order to use the GUI, you must first download the latest version, connect the PC to the SLICE-QTC, and assign a COM port to the SLICE-QTC. ​ The following describes the steps involved. It is possible to control the SLICE-QTC from a Windows<​sup>​®</​sup>​-based PC using a GUI.  In order to use the GUI, you must first download the latest version, connect the PC to the SLICE-QTC, and assign a COM port to the SLICE-QTC. ​ The following describes the steps involved.
  
-  - Download and store the latest GUI executable from the FTP site (Request ​copy of executable GUI from [[sales@vescent.com|sales [at] vescent [dot] com]]).+  - Download and store the latest GUI executable from [[ https://​github.com/​Vescent| ​the Vescent ​FTP site]] (or request ​copy of executable GUI from [[sales@vescent.com|sales [at] vescent [dot] com]]).
   - Connect the SLICE-QTC to the PC using a USB type B cable.   - Connect the SLICE-QTC to the PC using a USB type B cable.
   - Turn on the SLICE-QTC.   - Turn on the SLICE-QTC.