Line 51: Line 51:
 ===== Proper Usage ===== ===== Proper Usage =====
 | {{ :warning-sign.png?100&nolink }}  | If this instrument is used in a manner not specified by the manufacturer in this manual or other relevant literature, protection provided by the instrument may be impaired. | | {{ :warning-sign.png?100&nolink }}  | If this instrument is used in a manner not specified by the manufacturer in this manual or other relevant literature, protection provided by the instrument may be impaired. |
 +|:::| Successful implementation of the SLICE-QTC depends critically on the design of the whole system: controller, transducer, plant, and sensor. |
 +
 +==== Tips for Successful Temperature Regulation ==== 
 +  -  The transducer, plant, and sensor must all be in good thermal contact. Use thermal paste or epoxy to avoid air gaps between a TEC or heater and the body of the plant as well as to secure the thermistor to the plant.
 +  - Know the electrical limits of of your transducer and set the limits of the SLICE-QTC accordingly.  Each channel can provide up to  6 A or 20 W of power with a maximum compliance voltage up to ~18 V.  Choose transducers that work within these limits.  Using the current, power, and/or voltage ratings of your transducer to calculate loads on individual channels and set limits appropriately in the CH X > Settings > Load Limits window.
 +  - Maintain proper shielding of the control cable, but avoid ground loops between the SLICE-QTC and the plant.  In most cases this means extending the foil shield in the cable to as close as possible to the plant without electrically connecting the SLICE-QTC to the plant.  
  
 ===== List of Symbols ===== ===== List of Symbols =====
Line 71: Line 77:
 | Temperature Stability                                                                                                                     | ±0.2 mK over 1 hour((Plant-dependent; value specified for D2-100 laser))                                                                                                                                               | | Temperature Stability                                                                                                                     | ±0.2 mK over 1 hour((Plant-dependent; value specified for D2-100 laser))                                                                                                                                               |
 | Precision                                                                                                                                 | ~1 mK                                                                                                                                                                                                                  | | Precision                                                                                                                                 | ~1 mK                                                                                                                                                                                                                  |
-| Control Capacity                                                                                                                          | 40 W((User-distributable over four channels))                                                                                                                                                                          |+| Control Capacity                                                                                                                          | 40 W total((User-distributable over four channels)) \\ 20 W max for a single channel                                                                                                                                                                          |
 | Current Capacity                                                                                                                          | 6 A per channel((Subject to 40 W overall maximum))                                                                                                                                                                     | | Current Capacity                                                                                                                          | 6 A per channel((Subject to 40 W overall maximum))                                                                                                                                                                     |
 | Compliance Voltage                                                                                                                        | ≤18 V((automatically adjusted))                                                                                                                                                                                        | | Compliance Voltage                                                                                                                        | ≤18 V((automatically adjusted))                                                                                                                                                                                        |