ice:servo-opls [2019/11/18 23:30] – [Calculating Phase Noise] Michael Radunskyice:servo-opls [2019/11/18 23:31] – [Calculating Phase Noise] Michael Radunsky
Line 96: Line 96:
 The phase-noise specified in Section 1.3 is referenced to the phase frequency detector (PFD) at 1 Hz. To convert that to the noise measured on the actual beat-note, it must be rescaled with the following formula: The phase-noise specified in Section 1.3 is referenced to the phase frequency detector (PFD) at 1 Hz. To convert that to the noise measured on the actual beat-note, it must be rescaled with the following formula:
  
-<WRAP center round box 450px> **D2-135 Phase-Noise Floor = -213 + 20Log(N) + 10Log(F<sub>REF</sub>)**</WRAP>+<WRAP center round box 450px> **D2-135 Phase-Noise Floor = -213 + 20Log(N) + 10Log(ƒ<sub>REF</sub>)**</WRAP>
  
-where N is the value of the divider and F<sub>REF</sub> is the reference frequency as measured in Hz. For more details, please see this [[http://www.vescent.com/2012/calculating-phase-noise-from-the-d2-135/|application note]].+where N is the value of the divider and ƒ<sub>REF</sub> is the reference frequency as measured in Hz. For more details, please see this [[http://www.vescent.com/2012/calculating-phase-noise-from-the-d2-135/|application note]].
  
  
ice/servo-opls.txt · Last modified: 2021/08/26 15:26 by 127.0.0.1