User Tools

Site Tools


ice:quadtemp

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
ice:quadtemp [2014/07/22 20:11] – [I/O (OEM Only)] jtshugrueice:quadtemp [2021/08/26 15:26] (current) – external edit 127.0.0.1
Line 11: Line 11:
 ===== Description ===== ===== Description =====
  
-The Quad Temperature Controller Board (ICE-QT1) provides 4 independent PID temperature servo loops capable of sub-mK stability(( Sub-mK stability requires a proper thermal design and proper tuning of the temperature controller to the thermal plant. If you did not purchase the Laser Controller with a Laser Diode, please read the section on tuning the temperature controller.))Each channel can drive either a single TEC (bipolar current output) or resistive heater (unipolar current output) and read a single 10k thermistor. +The Quad Temperature Controller Board (ICE-QT1) provides 4 independent PID temperature servo loops capable of sub-mK stability.((The ICE-QT1 has non-adjustable poles.  They have been set appropriately for Vescent D2-100 DBR lasers and for Photodigm Mercury lasers. /* Sub-mK stability requires a proper thermal design and proper tuning of the temperature controller to the thermal plant. If you did not purchase the Laser Controller with a Laser Diode, please read the section on tuning the temperature controller. */)) Each channel can drive either a single TEC (bipolar current output) or resistive heater (unipolar current output) and read a single 10k thermistor. 
 ===== Absolute Maximum Ratings ===== ===== Absolute Maximum Ratings =====
 Note: All modules designed to be operated in laboratory environment Note: All modules designed to be operated in laboratory environment
Line 101: Line 101:
 </WRAP> </WRAP>
  
-The flat flex jumper cables used to connect to the FFC connectors should be 1.00mm pitch with 12 conductors. Another FFC connector must be used on the other side to break out the connections for OEM use. It should be noted that when connecting to the other side of the FFC, all the pin definitions will be reversed (i.e. pin 1 on one side will be pin 12 on the other). Generally with flat flex jumper cables, the cables can be folded but the bend radius should not be less that 2mm.+The flat flex jumper cables used to connect to the FFC connectors should be 1.00mm pitch with 12 conductors. Another FFC connector must be used on the other side to break out the connections for OEM use. It should be noted that when connecting to the other side of the FFC, all the pin definitions will be reversed (i.e. pin 1 on one side will be pin 12 on the other).  
 + 
 +<WRAP center round info 60%> 
 +Generally with flat flex jumper cables, the cables can be folded but the bend radius should not be less that 2mm. The clearance in front of the pcb FFC connectors must be at least 0.25 inches to allow the flat flex cable to exit and bend. The distance between the FFC connector and the first bend of the flat flex cable is determined by the length of the stiffener (or backing) on the flat flex cable. 
 +</WRAP> 
 ===== Quick Start Commands Guide ===== ===== Quick Start Commands Guide =====
  
ice/quadtemp.1406059870.txt.gz · Last modified: 2021/08/26 14:26 (external edit)