User Tools

Site Tools


ice:quadtemp

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
ice:quadtemp [2014/04/30 20:43] – external edit 127.0.0.1ice:quadtemp [2021/08/26 15:26] (current) – external edit 127.0.0.1
Line 11: Line 11:
 ===== Description ===== ===== Description =====
  
-The Quad Temperature Controller Board (ICE-QT1) provides 4 independent PID temperature servo loops capable of sub-mK stability(( Sub-mK stability requires a proper thermal design and proper tuning of the temperature controller to the thermal plant. If you did not purchase the Laser Controller with a Laser Diode, please read the section on tuning the temperature controller.))Each channel can drive either a single TEC (bipolar current output) or resistive heater (unipolar current output) and read a single 10k thermistor. +The Quad Temperature Controller Board (ICE-QT1) provides 4 independent PID temperature servo loops capable of sub-mK stability.((The ICE-QT1 has non-adjustable poles.  They have been set appropriately for Vescent D2-100 DBR lasers and for Photodigm Mercury lasers. /* Sub-mK stability requires a proper thermal design and proper tuning of the temperature controller to the thermal plant. If you did not purchase the Laser Controller with a Laser Diode, please read the section on tuning the temperature controller. */)) Each channel can drive either a single TEC (bipolar current output) or resistive heater (unipolar current output) and read a single 10k thermistor. 
 ===== Absolute Maximum Ratings ===== ===== Absolute Maximum Ratings =====
 Note: All modules designed to be operated in laboratory environment Note: All modules designed to be operated in laboratory environment
-<WRAP center round box 60%>+
 ^  Parameter  ^  Rating  ^ ^  Parameter  ^  Rating  ^
 |Environmental Temperature | >15°C and <30°C| |Environmental Temperature | >15°C and <30°C|
 |Environmental Humidity | <60% |  |Environmental Humidity | <60% | 
 |Environmental Dew Points | <15°C | |Environmental Dew Points | <15°C |
-</WRAP>+
  
 ===== Specifications  ===== ===== Specifications  =====
-<WRAP center round box 550px> +^ ^  Min  ^  Typical  ^  Max   Units  ^ 
-^ ^  ICE-QT1   Units  ^ +^ Temperature Overview^^^^^ 
-^ Temperature Overview^ ^ ^ +|Number of channels of Temperature Control    4  |  -  
-|<html> &nbsp;&nbsp; </html> Number of channels of Temperature Control  |  4  |   +|Loop Transfer Function((Not adjustable, fixed at factory))    PID  |  -  
-|<html> &nbsp;&nbsp; </html> Loop Transfer Function((Not adjustable, fixed at factory))  |  PID  |   +|Thermistor Resistance  |    10  |  -  |  kΩ  
-|<html> &nbsp;&nbsp; </html> Temperature Sensor  |  10kΩ thermistor    +^ Temperature Control^^^^^ 
-^ Temperature Control^ ^ ^ +|Temperature setpoint range |  -10     60  |  °C  | 
-|<html> &nbsp;&nbsp; </html> Temperature setpoint range |  -10  - 60  |  °C  | +|Temperature setpoint resolution |  |  0.1  |  °C  | 
-|<html> &nbsp;&nbsp; </html> Temperature setpoint resolution |  0.1  |  °C  | +|Temperature error resolution | | |  0.5  |  mK  | 
-|<html> &nbsp;&nbsp; </html> Temperature error resolution |  <0.5  |  mK  | +|Max  TEC current | | |  1  |  A  | 
-|<html> &nbsp;&nbsp; </html>Max  TEC current (voltage)|  1 (4)  |  A (V)  | +|Max  TEC voltage| | |  1   V  | 
-</WRAP>+ 
 +^Electrical Specifications^^^^^ 
 +^ ^  Min  ^  Typical  ^  Max  ^  Units  ^ 
 +|5V_A Current Draw((Maximum based on all four channels having TEC current limit set to maximum of 1 A. Typical represents normal stabilized current draw for two DBR lasers.)) | |  0.5  |  4   A  | 
 +|5V_D Current Draw | |  60  |    |  mA  | 
 +|+15V Current Draw | |  80  |    |  mA  | 
 +|-15V Current Draw | |  60  |    |  mA  | 
  
  
 ===== I/O (ICE-BOX)  ===== ===== I/O (ICE-BOX)  =====
  
-//Only when purchased with the ICE-BOX.//+<WRAP center round info 60%> 
 +Applies only when purchased with the [[ice:enclosure|ICE Platform]]. 
 +</WRAP>
  
  
Line 45: Line 54:
 The Front Panel for the ICE-QT1 has two 8-pin Hirose connector (HR25-7TR-8SA) that each carry the signals for two channels of temperature control.  The pin diagram below is for the top Hirose connector; the bottom connect has the same wiring but for channels 3 & 4 instead of 1 & 2. Rth and Rth_Rtn are the two ends of a 10k thermistor. The Front Panel for the ICE-QT1 has two 8-pin Hirose connector (HR25-7TR-8SA) that each carry the signals for two channels of temperature control.  The pin diagram below is for the top Hirose connector; the bottom connect has the same wiring but for channels 3 & 4 instead of 1 & 2. Rth and Rth_Rtn are the two ends of a 10k thermistor.
  
-<WRAP round center box 230px+<WRAP> 
-|  **Pin**   **Signal**  |+<tabcaption hirosePinoutHirose connecter pin definitions.> 
 + **Pin**   **Signal**  ^
 |  1  |  TEC1+  | |  1  |  TEC1+  |
 |  2  |  TEC1-  | |  2  |  TEC1-  |
Line 55: Line 65:
 |  7  |  Rth2  | |  7  |  Rth2  |
 |  8  |  Rth2-RTN  | |  8  |  Rth2-RTN  |
 +</tabcaption>
 </WRAP> </WRAP>
 +
  
 ===== I/O (OEM Only) ===== ===== I/O (OEM Only) =====
  
-//Only for OEM versions of the ICE-QT1 purchased without the ICE-Box//+<WRAP center round info 60%> 
 +Only for OEM versions of the ICE-QT1 purchased without the [[ice:enclosure|ICE Platform]]. 
 +</WRAP>
  
-**Laser Temp Output **+<imgcaption PCBDimensions|Component and connector positions on PCB.>{{ :ice:vpn00318_dimensions.jpg |}}</imgcaption> 
 + 
 +**Temperature Control Output Connections** 
 + 
 +The ICE-QT1 has two 12-pin 1.00mm pitch flat flex connectors ([[http://www.digikey.com/product-detail/en/SFW12R-1STE1LF/609-1900-1-ND/1003179|FCI SFW12R-1STE1LF]]) shown on the right side of <imgref PCBDimensions> that each carry the signals for two channels of temperature control.  The left-most connector carries signals for channels 1 & 2 while the right-most connector carries signals for channels 3 & 4. The pin diagram for the top flat flex connector (J2) is shown below and the bottom connector (J5) has the same pin definition for channels 3 & 4. Pin 1 is defined the same as noted in the datasheet for the [[http://portal.fciconnect.com/Comergent//fci/drawing/sfw12lf.pdf|SFW12R FFC connector]] and shown below in <imgref FFCpindef>. Rth and Rth_Rtn are the two ends of a 10k thermistor. The polarity for the TEC connections uses the assumption that current is flowing from the positive terminal to the negative, the TEC is heating.
  
-The ICE-QT1 has two 12-pin 1.00mm pitch flat flex connectors (FCI SFW12R-1STE1LFthat each carry the signals for two channels of temperature control The left-most connector carries signals for channels 1 & 2 while the right-most connector carries signals for channels 3 & 4. The pin diagram for the left-most flat flex connector is shown below and the right-most connector has the same pin definition for channels 3 & 4.  Rth and Rth_Rtn are the two ends of a 10k thermistor.+<imgcaption FFCpindef|Flat flex connector pin definition (pin 1 highlighted in yellow).>{{ :ice:sfw12r-1ste1lf_pin_diagram.jpg |}}</imgcaption>
  
-<WRAP round center box 230px+<WRAP> 
-|  **Pin**   **Signal**  |+<tabcaption ffcPinDefsFlat flex connector pin definitions> 
 + **Pin**   **Signal**  ^
 |  1  |  TEC1+  | |  1  |  TEC1+  |
 |  2  |  TEC1+  | |  2  |  TEC1+  |
Line 79: Line 98:
 |  11  |  Rth2  | |  11  |  Rth2  |
 |  12  |  Rth2-RTN  | |  12  |  Rth2-RTN  |
 +</tabcaption>
 +</WRAP>
 +
 +The flat flex jumper cables used to connect to the FFC connectors should be 1.00mm pitch with 12 conductors. Another FFC connector must be used on the other side to break out the connections for OEM use. It should be noted that when connecting to the other side of the FFC, all the pin definitions will be reversed (i.e. pin 1 on one side will be pin 12 on the other). 
 +
 +<WRAP center round info 60%>
 +Generally with flat flex jumper cables, the cables can be folded but the bend radius should not be less that 2mm. The clearance in front of the pcb FFC connectors must be at least 0.25 inches to allow the flat flex cable to exit and bend. The distance between the FFC connector and the first bend of the flat flex cable is determined by the length of the stiffener (or backing) on the flat flex cable.
 </WRAP> </WRAP>
  
ice/quadtemp.1398890610.txt.gz · Last modified: 2021/08/26 14:26 (external edit)