 List of common commands for controlling the laser controller in both the Peak Lock Servo and Peak Lock Servo & Current Controller and the Offset Phase Lock Servo & Current Controller. List of common commands for controlling the laser controller in both the Peak Lock Servo and Peak Lock Servo & Current Controller and the Offset Phase Lock Servo & Current Controller.
  
-<function=ICEfunc|0=Laser?&2=On&3=16>+<function=ICEfunc?0=Laser?&2=On&3=16>
 Returns the status of the laser (on, off or fault). Returns the status of the laser (on, off or fault).
 <html></div></html> <html></div></html>
  
-<function=ICEfunc|0=Laser&1=A:On/Off:On&2=Fault&3=17>+<function=ICEfunc?0=Laser&1=A:On/Off:On&2=Fault&3=17>
 Turns the laser on or off. Valid arguments are ON or OFF. Only turns the laser on when no fault / interlock issues. Returns the output from the Laser? command. Turns the laser on or off. Valid arguments are ON or OFF. Only turns the laser on when no fault / interlock issues. Returns the output from the Laser? command.
 <html></div></html> <html></div></html>
  
-<function=ICEfunc|0=CurrSet?&2=147.6&3=18>+<function=ICEfunc?0=CurrSet?&2=147.6&3=18>
 Returns the laser current setpoint (in mA) for how much current will flow through the laser when the laser is on. <html></div></html> Returns the laser current setpoint (in mA) for how much current will flow through the laser when the laser is on. <html></div></html>
  
-<function=ICEfunc|0=CurrSet&1=F:CURRENT:123.52&2=123.5&3=19>+<function=ICEfunc?0=CurrSet&1=F:CURRENT:123.52&2=123.5&3=19>
 Sets the setpoint laser current (in mA) to CURRENT. Returns the output from the **CurrSet?** command.  Sets the setpoint laser current (in mA) to CURRENT. Returns the output from the **CurrSet?** command. 
 <html></div></html> <html></div></html>
  
-<function=ICEfunc|0=CurrLim?&2=180.0&3=20>+<function=ICEfunc?0=CurrLim?&2=180.0&3=20>
 Returns the current limit (in mA) for the maximum current allowed to flow through the laser when the laser is on. Returns the current limit (in mA) for the maximum current allowed to flow through the laser when the laser is on.
 <html></div></html> <html></div></html>
  
-<function=ICEfunc|0=CurrLim&1=F:CURRENT:156&2=156&3=21>+<function=ICEfunc?0=CurrLim&1=F:CURRENT:156&2=156&3=21>
 Sets the maximum laser current to CURRENT. Returns the output from the **CurrLim?** command.  Sets the maximum laser current to CURRENT. Returns the output from the **CurrLim?** command. 
 //Note: If Laser Current is set above new current limit value, then current setpoint is lowered to new current limit value.// //Note: If Laser Current is set above new current limit value, then current setpoint is lowered to new current limit value.//
 <html></div></html> <html></div></html>
  