This is an old revision of the document!
Table of Contents
ICE Platform Enclosure
Document Revision: 1.0
Document Last Updated on 2021/08/26 14:26
Please read Limited Warranty and General Warnings and Cautions prior to operating.
Description
The ICE Platform Enclosure contains the necessary Master Control Board, communications ports, I/O, mounting hardware, and thermal management for supporting up to 8 daughter modules.
Absolute Maximum Ratings
Note: All modules designed to be operated in laboratory environment
Parameter | Rating |
---|---|
Environmental Temperature | >15°C and <30°C |
Environmental Humidity | <60% |
Environmental Dew Points | <15°C |
Specifications
ICE-MC1 | Units | |
---|---|---|
Input Power Requirements | ||
<html> </html> 5V line | >5 | A |
<html> </html> 15V line | >1.5 | A |
<html> </html> -15V line | >1.5 | A |
Maximum Power Consumption1) | ||
<html> </html> 5V line | 10 | A |
<html> </html> 15V line | 3 | A |
<html> </html> -15V line | 3 | A |
Slave Boards | ||
<html> </html> Max Number of Slave Boards | 8 | |
Communication Methods | ||
<html> </html> USB 2.0 | ||
<html> </html> TTL Serial (3.3V) |
I/O
USB
USB communication to the ICE-MC1 is done via a type-B USB connector on board. The ICE-MC1 emulates a serial port over USB and all communications with the device should be done with Serial communication settings of '8N1' (8 bit, no parity, 1 stop bit) at 115,200 baud. The comm (or serial) port to use will be chosen by the USB host computer.
TTL Serial
3.3V TTL Serial is available on the back 9pin D-sub connector on the ICE-Box. The pin definition is shown below:
Pin | Signal |
1 | RESERVED |
2 | RXI_Serial |
3 | TXO_Serial |
4 | 5V In2) |
5 | Ground |
6 | RESERVED |
7 | No Connect |
8 | RESERVED |
9 | RESERVED |
TXO is for an external device transmitting data out to the ICE-MC1. RXI is for the ICE-MC1 sending data that an external device receives.