d2:laser_servo [2019/08/28 23:26] – [Laser Lock Troubleshooting] Michael Radunskyd2:laser_servo [2019/09/06 17:35] – [Right Side Panel] Michael Radunsky
Line 271: Line 271:
 **Ramp Master / Slave (Jumper)** **Ramp Master / Slave (Jumper)**
  
-This jumper is only accessible by removing the right side panel (see above) and sets whether the ramp input is in master or slave mode. It is factory set to be in MASTER MODE. In SLAVE MODE (jumper off) the RAMP signal is generated externally and input through the back panel RAMP I/O port. In MASTER MODE (jumper on) the ramp is generated internally and is sent out to the RAMP I/O port for driving other D2-125 Laser Servos configured in SLAVE MODE.+This jumper is only accessible by removing the right side panel (see above) and sets whether the ramp input is in master or slave mode. It is factory set to be in MASTER MODE. In SLAVE MODE (jumper off) the RAMP signal is generated externally and input through the back panel RAMP I/O port. In MASTER MODE (jumper on) the ramp is generated internally and is sent out to the RAMP I/O port for driving other D2-125 Laser Servos configured in SLAVE MODE.   
 + 
 +The amplitude of the slave ramp is about -5x of the master ramp input (at slave Ramp Amp maximum).  The DC offset is applied after amplification.
  
 ====Lock Guard==== ====Lock Guard====
d2/laser_servo.txt · Last modified: 2021/12/17 01:58 by 127.0.0.1