d2:laser_servo [2019/02/21 15:55] – [Right Side Panel] Michael Radunskyd2:laser_servo [2019/07/15 23:00] – [Laser Lock Troubleshooting] Michael Radunsky
Line 185: Line 185:
 **Auxiliary Servo Output** **Auxiliary Servo Output**
  
-The AUXILIARY SERVO OUTPUT is generated from integrating the SERVO OUTPUT.  Its purpose is to supply a correction signal to drive the SERVO OUTPUT to zero. When used with Vescent DBR lasers and the D2-105 Laser Controller, the AUXILIARY SERVO OUTPUT can be connected to the TEMP SERVO IN to adjust the laser diode temperature to keep the feedback laser current constant. Similarly, AUXILIARY SERVO OUTPUT can drive a PZT on an external-cavity laser diode to keep the laser diode current constant. See [[d2:laser_servo#Aux Gain: + / Aux Gain - (2-position switch)|AUXILIARY SERVO: GAIN SIGN]] and [[d2:laser_servo#Auxiliary Servo: Gain (25-turn trimpot)|AUXILIARY SERVO: GAIN]] for information on setting the gain and and gain sign of the AUXILIARY SERVO OUTPUT.+The AUXILIARY SERVO OUTPUT is generated from integrating the SERVO OUTPUT.  Its purpose is to supply a correction signal to drive the SERVO OUTPUT to zero. When used with Vescent DBR lasers and the D2-105 Laser Controller, the AUXILIARY SERVO OUTPUT can be connected to the TEMP SERVO IN to adjust the laser diode temperature to keep the feedback laser current constant. Similarly, AUXILIARY SERVO OUTPUT can drive a PZT on an external-cavity laser diode to keep the laser diode current constant. See [[d2:laser_servo#Aux Gain: + / Aux Gain - (2-position switch)|AUXILIARY SERVO: GAIN SIGN]] and [[d2:laser_servo#Auxiliary Servo: Gain (25-turn trimpot)|AUXILIARY SERVO: GAIN]] for information on setting the gain and gain sign of the AUXILIARY SERVO OUTPUT.
  
 **Ramp Offset**  **Ramp Offset** 
Line 243: Line 243:
 ===Auxiliary Servo: Gain (25-turn trimpot)=== ===Auxiliary Servo: Gain (25-turn trimpot)===
  
-This trimpot sets the gain for the AUXILIARY SERVO OUTPUT. The Aux Servo is a pure integrator and the gain ranges from 500 μs to 5 s.+This trimpot sets the gain for the AUXILIARY SERVO OUTPUT. The Aux Servo is a pure integrator and the gain ranges from 500 μs to 5 s.  Full ccw sets gain to minimum (5 s), full cw sets gain to maximum (500 µs).
  
 <WRAP center round box 420px><imgcaption d2_125_side_panel_photo|Certain rarely used controls are only accessible by removing the right-side panel, as shown in the figure above.> <WRAP center round box 420px><imgcaption d2_125_side_panel_photo|Certain rarely used controls are only accessible by removing the right-side panel, as shown in the figure above.>
Line 375: Line 375:
 If you are having problems locking the laser, it is a good idea to not use the AUXILIARY SERVO OUTPUT as this complicates the system. Once you get the locking to work properly, you can reconnect this cable. If you are having problems locking the laser, it is a good idea to not use the AUXILIARY SERVO OUTPUT as this complicates the system. Once you get the locking to work properly, you can reconnect this cable.
  
-<imgref factory_settings> shows the nominal corner settings for locking the Vescent D2-100 to atomic spectroscopy.  If all else fails, return to these setting, turn down the gain and try again.+<imgref factory_settings> shows and <tabref factory_setting> lists the nominal corner settings for locking the Vescent D2-100 (or a Photodigm TOSA) to atomic spectroscopy.  If all else fails, return to these setting, turn down the gain and try again.
  
 <WRAP center round box 400px><imgcaption factory_settings|Factory settings for locking a D2-100> <WRAP center round box 400px><imgcaption factory_settings|Factory settings for locking a D2-100>
-{{ {{ d2:d2-125:factory_settings.jpg?direct&400 |}} |}}</imgcaption></WRAP>+{{ {{ d2:d2-125:factory_settings_red.png?direct&400 |}} |}}</imgcaption></WRAP> 
 + 
 +<WRAP center round box 60%><tabcaption factory_setting |D2-125 locking D2-100 or TOSA> 
 +| **Corner**                                                       | **Value**       | **Units** 
 +| First Integrator (high freq.)                                    | low freq.             | 
 +| First Integrator (low freq.)                                     | 10              | Hz         | 
 +| Second Integrator (high freq.)                                   | low freq.             | 
 +| Second Integrator (low freq.)                                    | 500             | Hz         | 
 +| Differential (high freq.)                                        | OFF             | 
 +| Differential (low freq.)                                         | 100              | kHz         | 
 + 
 +</tabcaption></WRAP>
  