d2:laser_servo [2019/01/29 21:56] – [Back-panel Section] Michael Radunskyd2:laser_servo [2019/07/15 22:50] – [Laser Lock Troubleshooting] Michael Radunsky
Line 185: Line 185:
 **Auxiliary Servo Output** **Auxiliary Servo Output**
  
-The AUXILIARY SERVO OUTPUT is generated from integrating the SERVO OUTPUT.  Its purpose is to supply a correction signal to drive the SERVO OUTPUT to zero. When used with Vescent DBR lasers and the D2-105 Laser Controller, the AUXILIARY SERVO OUTPUT can be connected to the TEMP SERVO IN to adjust the laser diode temperature to keep the feedback laser current constant. Similarly, AUXILIARY SERVO OUTPUT can drive a PZT on an external-cavity laser diode to keep the laser diode current constant. See [[d2:laser_servo#Aux Gain: + / Aux Gain - (2-position switch)|AUXILIARY SERVO: GAIN SIGN]] and [[d2:laser_servo#Auxiliary Servo: Gain (25-turn trimpot)|AUXILIARY SERVO: GAIN]] for information on setting the gain and and gain sign of the AUXILIARY SERVO OUTPUT.+The AUXILIARY SERVO OUTPUT is generated from integrating the SERVO OUTPUT.  Its purpose is to supply a correction signal to drive the SERVO OUTPUT to zero. When used with Vescent DBR lasers and the D2-105 Laser Controller, the AUXILIARY SERVO OUTPUT can be connected to the TEMP SERVO IN to adjust the laser diode temperature to keep the feedback laser current constant. Similarly, AUXILIARY SERVO OUTPUT can drive a PZT on an external-cavity laser diode to keep the laser diode current constant. See [[d2:laser_servo#Aux Gain: + / Aux Gain - (2-position switch)|AUXILIARY SERVO: GAIN SIGN]] and [[d2:laser_servo#Auxiliary Servo: Gain (25-turn trimpot)|AUXILIARY SERVO: GAIN]] for information on setting the gain and gain sign of the AUXILIARY SERVO OUTPUT.
  
 **Ramp Offset**  **Ramp Offset** 
Line 243: Line 243:
 ===Auxiliary Servo: Gain (25-turn trimpot)=== ===Auxiliary Servo: Gain (25-turn trimpot)===
  
-This trimpot sets the gain for the AUXILIARY SERVO OUTPUT. The Aux Servo is a pure integrator and the gain ranges from 500 μs to 5 s.+This trimpot sets the gain for the AUXILIARY SERVO OUTPUT. The Aux Servo is a pure integrator and the gain ranges from 500 μs to 5 s.  Full ccw sets gain to minimum (5 s), full cw sets gain to maximum (500 µs).
  
 <WRAP center round box 420px><imgcaption d2_125_side_panel_photo|Certain rarely used controls are only accessible by removing the right-side panel, as shown in the figure above.> <WRAP center round box 420px><imgcaption d2_125_side_panel_photo|Certain rarely used controls are only accessible by removing the right-side panel, as shown in the figure above.>
Line 255: Line 255:
 ** Ramp->Aux / Ramp->Servo (2-position slider switch) ** ** Ramp->Aux / Ramp->Servo (2-position slider switch) **
  
-This 2-position slider switch is only accessible by removing the right side panel (see above) and sets whether the ramp is applied to the SERVO OUTPUT or the AUXILIARY SERVO OUTPUT. It is factory set in the Ramp->Servo position. When the Ramp->Servo is selected, the ramp is a ~500Hz triangle wave and when Ramp->Aux is selected the ramp is ~50Hz. Additionally, when using The Absolute Jump TTL or the Relative Jump TTL, the output jump is placed on the same channel as gets the ramp signal. The other signal (SERVO OUTPUT or AUXILIARY SERVO OUTPUT) is held at its current value. +This 2-position slider switch is only accessible by removing the right side panel (see above) and sets whether the ramp is applied to the SERVO OUTPUT or the AUXILIARY SERVO OUTPUT. It is factory set in the Ramp->Servo position. When the Ramp->Servo is selected, the ramp is a ~500 Hz triangle wave and when Ramp->Aux is selected the ramp is ~50 Hz. Additionally, when using The Absolute Jump TTL or the Relative Jump TTL, the output jump is placed on the same channel as gets the ramp signal. The other signal (SERVO OUTPUT or AUXILIARY SERVO OUTPUT) is held at its current value. 
  
 ===Aux Gain: + / Aux Gain - (2-position switch)=== ===Aux Gain: + / Aux Gain - (2-position switch)===
Line 378: Line 378:
  
 <WRAP center round box 400px><imgcaption factory_settings|Factory settings for locking a D2-100> <WRAP center round box 400px><imgcaption factory_settings|Factory settings for locking a D2-100>
-{{ {{ d2:d2-125:factory_settings.jpg?direct&400 |}} |}}</imgcaption></WRAP>+{{ {{ d2:d2-125:factory_settings_red.png?direct&400 |}} |}}</imgcaption></WRAP>
  
  =====D2-125 Stand-alone Test=====  =====D2-125 Stand-alone Test=====