User Tools

Site Tools


d2:laser_controller

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
Next revisionBoth sides next revision
d2:laser_controller [2021/06/10 21:20] – external edit 127.0.0.1d2:laser_controller [2021/12/17 00:00] – external edit 127.0.0.1
Line 254: Line 254:
 {{ :d2:d-sub_power_pinout.jpg?nolink&150 |}} {{ :d2:d-sub_power_pinout.jpg?nolink&150 |}}
  
 +While it is infrequent, the D2-005 power supply may occasionally radiate noise from the side of its chassis onto nearby electronics. This only occurs in some system configurations, and will appear as a signal at the frequency of your mains electricity (typically either 50 Hz or 60 Hz). This noise can easily be removed by moving the D2-005 at least 18 inches (45cm) away from other electronics, rotating it 90° such that the sides of the D2-005 face away, or by moving the entire power supply to a different shelf. To accommodate this, all D2-005's are shipped with a 5' DB9 cable as of January 1, 2022. 
  
 **Laser Enable** **Laser Enable**
Line 307: Line 308:
   - Measure the resistance across GND5 and probe point GAIN2. Set this resistance to between 1.2 kΩ and 1.4 kΩ by adjusting trimpot PROPGAIN2.   - Measure the resistance across GND5 and probe point GAIN2. Set this resistance to between 1.2 kΩ and 1.4 kΩ by adjusting trimpot PROPGAIN2.
  
-<WRAP center round box 570px><imgcaption factory_set_post6987|Factory settings of T2 (left) and T1 (right) poles for units with S/N 2675 or higher (PCBs with S/N 6987 and higher)>[{{ :d2:d2-105:temp_loop_post_sn_6987_w_text.png?nolink&550 | filler text }}]</imgcaption></WRAP>+<WRAP center round box 570px><imgcaption factory_set_post6987|Factory settings of T2 (left) and T1 (right) poles for units with S/N 2675 or higher (PCBs with S/N 6987 and higher)>[{{ :d2:d2-105:Modern_Corners.jpg?nolink&550 | filler text }}]</imgcaption></WRAP>
  
 <WRAP center round box 60%><tabcaption factory_set2_post6987|Factory settings of T2 Proportional poles (first DIP switch in figure) and T2 Differential poles (second DIP switch in figure) for units with S/N 2675 or higher (PCBs with S/N 6987 and higher)> <WRAP center round box 60%><tabcaption factory_set2_post6987|Factory settings of T2 Proportional poles (first DIP switch in figure) and T2 Differential poles (second DIP switch in figure) for units with S/N 2675 or higher (PCBs with S/N 6987 and higher)>
d2/laser_controller.txt · Last modified: 2024/03/27 15:33 by Thomas Bersano