Differences
This shows you the differences between two versions of the page.
Both sides previous revisionPrevious revisionNext revision | Previous revisionNext revisionBoth sides next revision |
d2:laser_controller [2021/06/10 21:20] – external edit 127.0.0.1 | d2:laser_controller [2021/12/17 00:00] – external edit 127.0.0.1 |
---|
{{ :d2:d-sub_power_pinout.jpg?nolink&150 |}} | {{ :d2:d-sub_power_pinout.jpg?nolink&150 |}} |
| |
| While it is infrequent, the D2-005 power supply may occasionally radiate noise from the side of its chassis onto nearby electronics. This only occurs in some system configurations, and will appear as a signal at the frequency of your mains electricity (typically either 50 Hz or 60 Hz). This noise can easily be removed by moving the D2-005 at least 18 inches (45cm) away from other electronics, rotating it 90° such that the sides of the D2-005 face away, or by moving the entire power supply to a different shelf. To accommodate this, all D2-005's are shipped with a 5' DB9 cable as of January 1, 2022. |
| |
**Laser Enable** | **Laser Enable** |
- Measure the resistance across GND5 and probe point GAIN2. Set this resistance to between 1.2 kΩ and 1.4 kΩ by adjusting trimpot PROPGAIN2. | - Measure the resistance across GND5 and probe point GAIN2. Set this resistance to between 1.2 kΩ and 1.4 kΩ by adjusting trimpot PROPGAIN2. |
| |
<WRAP center round box 570px><imgcaption factory_set_post6987|Factory settings of T2 (left) and T1 (right) poles for units with S/N 2675 or higher (PCBs with S/N 6987 and higher)>[{{ :d2:d2-105:temp_loop_post_sn_6987_w_text.png?nolink&550 | filler text }}]</imgcaption></WRAP> | <WRAP center round box 570px><imgcaption factory_set_post6987|Factory settings of T2 (left) and T1 (right) poles for units with S/N 2675 or higher (PCBs with S/N 6987 and higher)>[{{ :d2:d2-105:Modern_Corners.jpg?nolink&550 | filler text }}]</imgcaption></WRAP> |
| |
<WRAP center round box 60%><tabcaption factory_set2_post6987|Factory settings of T2 Proportional poles (first DIP switch in figure) and T2 Differential poles (second DIP switch in figure) for units with S/N 2675 or higher (PCBs with S/N 6987 and higher)> | <WRAP center round box 60%><tabcaption factory_set2_post6987|Factory settings of T2 Proportional poles (first DIP switch in figure) and T2 Differential poles (second DIP switch in figure) for units with S/N 2675 or higher (PCBs with S/N 6987 and higher)> |