Line 326: | Line 326: | ||
To reset the temperature loop parameters to the approximate settings for a Photodigm TOSA DBR Laser: | To reset the temperature loop parameters to the approximate settings for a Photodigm TOSA DBR Laser: | ||
- The T1 loop is not used as the TOSA only has one TEC. | - The T1 loop is not used as the TOSA only has one TEC. | ||
- | - Set the dip switches to the configuration shown in <tabref factory_set2_post6987_TOSA> | + | - Set the T2 dip switches to the configuration shown in <tabref factory_set2_post6987_TOSA> |
- Measure the resistance across GND5 and probe point GAIN2. Set this resistance to between 1.01 kΩ and 1.03 kΩ by adjusting trimpot PROPGAIN2. | - Measure the resistance across GND5 and probe point GAIN2. Set this resistance to between 1.01 kΩ and 1.03 kΩ by adjusting trimpot PROPGAIN2. | ||