Line 326: Line 326:
 To reset the temperature loop parameters to the approximate settings for a Photodigm TOSA DBR Laser: To reset the temperature loop parameters to the approximate settings for a Photodigm TOSA DBR Laser:
   - The T1 loop is not used as the TOSA only has one TEC.   - The T1 loop is not used as the TOSA only has one TEC.
-  - Set the dip switches to the configuration shown in <tabref factory_set2_post6987_TOSA>+  - Set the T2 dip switches to the configuration shown in <tabref factory_set2_post6987_TOSA>
   - Measure the resistance across GND5 and probe point GAIN2. Set this resistance to between 1.01 kΩ and 1.03 kΩ by adjusting trimpot PROPGAIN2.   - Measure the resistance across GND5 and probe point GAIN2. Set this resistance to between 1.01 kΩ and 1.03 kΩ by adjusting trimpot PROPGAIN2.
  