Line 20: | Line 20: | ||
=====Purchase Includes===== | =====Purchase Includes===== | ||
* D2-250 Heterodyne Module | * D2-250 Heterodyne Module | ||
+ | * VPN00460 Fiber Optic Cable (multimode) | ||
+ | * F220FC-B Fiber Coupler | ||
+ | * 5/64" Hex Key | ||
===== Absolute Maximum Ratings ===== | ===== Absolute Maximum Ratings ===== | ||
Line 66: | Line 69: | ||
===Procedure=== | ===Procedure=== | ||
Mount both laser modules and the D2-250 Heterodyne Module to an optical breadboard or table. Leave sufficient space between the lasers and optical modules to allow for the adjustment of the hex screws and insertion of the F220FC-B Fiber Coupler and VPN00460 Fiber Optic Cable. Turn on both lasers. During mounting, align the first laser such that the beam enters " | Mount both laser modules and the D2-250 Heterodyne Module to an optical breadboard or table. Leave sufficient space between the lasers and optical modules to allow for the adjustment of the hex screws and insertion of the F220FC-B Fiber Coupler and VPN00460 Fiber Optic Cable. Turn on both lasers. During mounting, align the first laser such that the beam enters " | ||
+ | |||
+ | < | ||
Turn OFF both lasers, then screw the F220FC-B Fiber Coupler into the "FC Port" on the D2-250 and hand tighten. Screw in the FC/APC connector of the VPN00460 Fiber Optic Cable into the F220FC-B Fiber Coupler and hand tighten. | Turn OFF both lasers, then screw the F220FC-B Fiber Coupler into the "FC Port" on the D2-250 and hand tighten. Screw in the FC/APC connector of the VPN00460 Fiber Optic Cable into the F220FC-B Fiber Coupler and hand tighten. | ||
- | With the power meter, observe the output power through the VPN00460 Fiber Optic Cable. With the 5/64" hex key, adjust the first kinematic mount in the " | + | With the power meter, observe the output power through the VPN00460 Fiber Optic Cable. With the 5/64" hex key, adjust the first kinematic mount in the " |
+ | |||
+ | < | ||
Fully optimize the coupling into the fiber by adjusting the " | Fully optimize the coupling into the fiber by adjusting the " |